Overview
This technical note describes the procedure for scaling a Red Lion PAXS0000 Strain Gage Input Meter to display engineering units (weight) from a load cell. The procedure uses the meter's two-point scaling feature to correlate the load cell's mV output to the actual weight applied to the scale.
Applicable Products
- PAXS0000 Strain Gage Input Meter
Prerequisites
Before beginning, gather the following information:
- Load cell sensitivity (mV/V rating), typically listed on the load cell nameplate or datasheet.
- A known calibration weight or weights.
- Total weight value that will be applied during calibration.
- Desired display resolution (decimal point placement).
Configuration Procedure
1. Enter the Input Configuration Menu
- Press PAR until Pro No is displayed.
- Press F1 once to access 1-INP.
- Press PAR to display RANGE.
2. Configure the Load Cell Range
RANGE = ?.?A
Enter the load cell mV/V sensitivity multiplied by 10.
Examples:
| Load Cell Rating | RANGE Setting |
|---|---|
| 2.0 mV/V | 20.0A |
| 3.0 mV/V | 30.0A |
| 3.5 mV/V | 35.0A |
3. Configure Basic Display Parameters
Press PAR through the following parameters:
| Parameter | Setting |
|---|---|
| DECPT | Desired decimal point location |
| ROUND | Default |
| FILTR | Default |
| BAND | Default |
| PTS | 2 |
4. Configure the Scaling Method
- Press PAR until STYLE appears.
- Press F1.
- Change the setting from KEY to APLY.
- Press PAR.
5. Capture the Zero Scale Point
INP-1
With no weight applied to the scale:
- Observe the value displayed for INP-1.
- Record the value.
DSP-1
Set:
DSP-1 = 0
This establishes the zero-weight reference point.
6. Capture the Span (Loaded) Scale Point
INP-2
- Place the known calibration weight on the scale.
- Allow the reading to stabilize.
- Record the value displayed for INP-2.
DSP-2
Enter the actual weight applied to the scale.
Example:
- Calibration weight applied = 500 lb
- Set:
DSP-2 = 500
or
DSP-2 = 500.0
depending on the selected decimal point configuration.
7. Save the Configuration
- Press PAR until Pro No is displayed.
- Continue pressing PAR until END appears.
- The meter will return to normal operation with the new scaling applied.
Example Configuration
Load Cell:
- Sensitivity: 3.0 mV/V
Calibration Weight:
- 1,000 lb
Settings:
| Parameter | Value |
|---|---|
| RANGE | 30.0A |
| DECPT | 0 |
| PTS | 2 |
| STYLE | APLY |
| DSP-1 | 0 |
| DSP-2 | 1000 |
When complete, the PAXS0000 will display weight directly in pounds based on the load cell input signal.
Troubleshooting
| Symptom | Possible Cause |
|---|---|
| Display reads incorrectly at all weights | Incorrect RANGE value entered |
| Display does not return to zero when unloaded | Load cell not mechanically unloaded or INP-1 captured incorrectly |
| Reading is unstable | Electrical noise, grounding issue, or load cell wiring problem |
| Full-scale reading is incorrect | Wrong calibration weight entered in DSP-2 or incorrect INP-2 value recorded |
| Reading changes after calibration | Load cell capacity or excitation wiring does not match the application |
Notes
- The calibration weight should be as close as possible to the normal operating range of the scale for best accuracy.
- Higher calibration weights generally provide better scaling accuracy than very small test weights.
- If multiple load cells are connected through a summing box, ensure the system is properly balanced before performing calibration.
- The PAXS0000 uses linear two-point scaling between the unloaded and loaded reference points established during this procedure.